The AvaSpec-2048x14 Fiber Optic Spectrometer is a back-thinned type CCD spectrometer with high quantum efficiency and high UV sensitivity. The optical design is based on the AvaBench-75 symmetrical Czerny-Turner design with a 2048x14 pixels high UV sensitive CCD image sensor. The image sensor is used as a linear array of 2048 pixels binning the vertical 14 pixels to optimize efficiency. The spectrometer has a fiber optic entrance connector (standard SMA, others possible), collimating and focusing mirror and a diffractional grating. A choice of 16 different gratings with different dispersion and blaze angles enable applications in the 200-1160nm range. The AvaSpec-2048x14 comes with a 16 bit AD converter, and USB2.0 high speed interface. The AvaSpec- 2048x14 is especially suitable for measuring low light, fluorescence  and UV- applications. Digital IO ports enable external triggering and control of shutter and pulsed light sources from the Avantes line of instruments.

The AvaSpec-2048x14 has a USB2 interface with fast data sampling of 450 spectra per second and data transfer in 2.24 ms and supports analog in- and outputs as well. Optional Bluetooth® (-BT) communication and an SD card for on-board saving of spectra can be added. The AvaSpec-2048x14-USB2 runs on USB power and comes with AvaSoft-basic, a complete manual and USB interface cable. Multiple (up to 127) USB2 spectrometers with different detector types can be externally coupled (see section multi-channel spectrometers, page 37). 


Technical Data
Optical Bench
Symmetrical Czerny-Turner, 75 mm focal length
Wavelength range
200 - 1160nm
Resolution
0.04 -20 nm, depending on configuration (see table)
Stray light
< 0.1%
Sensitivity (AvaLight-HAL, 8 µm fiber)
16.000 counts (16-bit AD)/µW -per ms integration time
UV Quantum efficiency
35-65% (200-300nm)
Detector
Back-thinned CCD image sensor 2048x14 pixels
Signal/Noise
500:1
AD converter
16 bit, 1.5 MHz
Integration time
2.24 ms - 10 minutes
Interface
USB 2.0 high speed, 480 Mbps
RS-232, 115.200 bps
Sample speed with on-board averaging
2.24 ms /scan
Data transfer speed
2.24 ms / scan (USB2)
432 ms / scan (RS-232)
Digital IO
HD-26 connector, 2 Analog in, 2 Analog out, 3 Digital in, 12 Digital out, trigger, synchronization
Power supply
Default USB power, 350 mA
Or with SPU2 external 12VDC, 350 mA
Dimensions, weight
175 x 110 x 44 mm(1 channel), 716 grams

Grating selection table for AvaSpec-2048x14

Use
Useable range 
Spectral range (nm)
Lines/mm
Blaze (nm)
Order code
UV/VIS/NIR
200-1160**
900**
300
300
UA
UV/VIS
200-850
520
600
300
UB
UV
200-750
250-220*
1200
250
UC
UV
200-650
165-145*
1800
UV
UD
UV
200-580
115-70*
2400
UV
UE
UV
220-400
70-45*
3600
UV
UF
UV/VIS
250-850
520
600
400
BB
VIS/NIR
300-1160**
800**
300
500
VA
VIS
360-1000
500
600
500
VB
VIS
300-800
250-200*
1200
500
VC
VIS
350-750
145-100*
1800
500
VD

VIS

350-640
75-50*
2400
VIS
VE
NIR
500-1050
500
600
750
NB
NIR
500-1050
220-150*
1200
750
NC
NIR
600-1160**
500**
300
1000
IA
NIR
600-1160
500
600
1000
IB

* depends on the starting wavelength of the grating; the higher the wavelength, the bigger the dispersion and the smaller the range to select.
**  please note that not all 2048 pixels will be used for the useable range  

Resolution table (FWHM) for AvaSpec-2048x14

Slit size (µm)
Grating (lines/mm)
10
25
50
100
200
500
300
0.8
1.4
2.4
4.3
8.0
20.0
600
0.4
0.7
1.2
2.1
4.1
10.0
1200
0.1-0.2*
0.2-0.3*
0.4-0.6*
0.7-1.0*
1.4-2.0*
3.3-4.8*
1800
0.07-0.12*
0.12-0.21*
0.2-0.36*
0.4-0.7*
0.7-1.4*
1.7-3.3*
2400
0.05-0.09*
0.08-0.15*
0.14-0.25*
0.3-0.5*
0.5-0.9*
1.2-2.2*
3600
0.04-0.06*
0.07-0.10*
0.11-0.16*
0.2-0.3*
0.4-0.6*
0.9-1.4*


 

 

 

 


*depends on the starting wavelength of the grating; the higher the wavelength, the bigger the dispersion and the better the resolution


Ordering Information
AvaSpec-2048x14-USB2 Fiber Optic Spectrometer, 75 mm Avabench, 2048x14 pixel back-thinned CCD detector, USB powered high speed USB2 interface, incl. AvaSoft-Basic, USB interface cable, specify grating, wavelength range and options
   
Options  
-SPU2 Incl. switch for USB2 power or external power for RS-232/BT
-SPU2-BT Bluetooth® interface for USB2 platform only, including antenna and switch
SDXXX  Internal XXX MB SD card for on-board data saving
DCL-UV/VIS-200 Quartz Detector Collection Lens (200-1100nm) 
SLIT-XX   Slit size, please specify XX = 10, 25, 50, 100, 200, 500 µm
OSF-YYY-3   Order sorting filter for reduction of 2nd order effects, 3 mm thick, please specify YYY= 385, 475, 515, 550, 600 nm
OSC Order sorting coating with 590nm long pass filter for VA, BB (>350nm) and VB gratings in AvaSpec-2048x14
OSC-UA Order sorting coating with 350 and 590nm longpass filter for UA gratings in AvaSpec-2048
OSC-UB Order sorting coating with 350 and 590nm longpass filter for UB or BB (<350nm) gratings in AvaSpec-2048x14

 







more categories

UV / VIS Spectrometers

NIR spectrometers

OEM spectrometers