home arrow Products arrow Spectrometers arrow Spectrometer Systems arrow AvaLIBS-50 and AvaLIBS-100 Laser Induced Breakdown Spectroscopy





AvaLIBS-50 and AvaLIBS-100 Laser Induced Breakdown Spectroscopy



AvaLIBS is a novel-design modular system which utilises Laser-Induced Breakdown Spectroscopy (LIBS) for the elemental characterisation of materials.  The highly modular and versatile design adopted in AvaLIBS will appeal to scientists and engineers interested in developing LIBS technology for their particular application area.  AvaLIBS is also well-suited for use in an academic teaching laboratory where ease of use and safety are of paramount importance.

AvaLIBS may be operated in “open-beam” mode as a Class IV laser system or fitted with one of a range of fully-interlocked modular sample chambers as a Class I laser system.  In either configuration the system can be operated with the AvaLIBS Head held or mounted in any orientation offering further flexibility in the types of sample that can be analysed. 

Features

- Modular and versatile design, suitable for laboratory and field applications

 - High-efficiency plasma light collection optics (3 UV-Vis channels and 3 Vis-NIR channels)

- Laser options: 50 mJ 1064 nm laser (AvaLIBS-50) and 100 mJ 1064 nm laser (AvaLIBS-100)

- Gas purge feature (for connection to external inert gas supply – Argon, Nitrogen, Helium)

- Optional imaging camera for viewing the sample before, during and after LIBS analysis

- May be operated with one of a number of removable, modular sample chambers (to Class I laser safety standards) or without (Class IV “open beam” configuration)

- Optional vertical support stand with height adjustment (facilitates operation of AvaLIBS Head in vertical orientation – with or without sample chamber)

- Up to six spectrometer modules may be installed (approx. 185 – 900 nm)

- External laser power supply easily disconnected to facilitate transportation

- May be operated with laptop PC or other suitable device running Microsoft Windows

 

 libscan_open_beam_next_to_metal_stock.jpg  0168-display_of_01_and_03_1.jpg
    
libscan_vertical_with_powder_sample_-_view_6.jpg
 0168-08-01a_1.jpg 0168-optical_table_ga_1.jpg 

 

 


System Specifications AvaLIBS base unit

AvaLIBS Head
Integrated laser beam expander and six-channel plasma light collection optics.  Horizontal or vertical operation, connected to Spectrometer console by 1.7 m flexible umbilical containing 6 optical fibers.Approx. dimensions  300x130x120 mm, weight 4 kg
Software
Specline and AvaSoft
AvaLIBS Spectrometer console(up to 6 spectrometer channels)
Contains spectrometers (up to 6), sync cables, USB hub, gas purge inlet (feeds purge gas to AvaLIBS head), laser safety interlock circuit, trigger input to spectrometers.Approx. dimensions  335 x 375 x 160 mm, weight 8 kg
Computer Interface
USB 2.0
Power connections
5VDC, 3A via plug-in adapter 100-240VAC(included)


Laser Specifications

 
YAG-LASER-50
YAG-LASER-100
Laser Type
Quantel Big Sky UltraQ-switched Nd: YAG (Class 4)
Quantel Big Sky UltraQ-switched Nd: YAG (Class 4)
Wavelength
1064 nm
1064 nm
Energy/Pulse
50 mJ/pulse
100 mJ/pulse
Pulse duration
7 ns (±2ns)
7 ns (±2ns)
Beam diameter
3 mm
4 mm
Repetition Rate
Up to 20 Hz
Up to 20 Hz
Power Supply
110-220 VAC, 800W
110-220 VAC, 800W
Power supply approx. dimensions, weight
360 mm x 435 mm x 140 mm, 14 kg
360 mm x 435 mm x 140 mm, 14 kg
 

Spectrometer Specifications

 

Spectrometer
AvaSpec-2048-LIBS1AvaSpec-2048-LIBS2AvaSpec-2048-LIBS3
Wavelength range
(185-256nm), FWHM 0.1nm(255-315nm), FWHM 0.1nm(314-316nm), FWHM 0.2nm
Spectrometer
AvaSpec-2048-LIBS4AvaSpec-2048-LIBS5AvaSpec-2048-LIBS6
Wavelength range
(414-498nm), FWHM 0.2nm(496-718nm), FWHM 0.5nm(716-904nm), FWHM 0.4nm
Resolution possible
< 0.1 nm
Detector type
CCD, 2048 pixels per channel
Programmable Integration time delay
min 1.28 µsec, steps 21 ns
Integration time
Min. 1.1 ms


AvaSoft operating software

The AvaSoft software controls the exact timing between the laser pulse and the start of the integration time. The timing is controlled by the laser or external pulse generator and synchronized with the AvaSpec spectrometer by running the spectrometer in external trigger mode. In this mode, the spectrometer receives a TTL-input signal from the laser, or simultaneously with the laser. The integration time delay is synchronized with this input signal.The AvaSoft software allows averaging, setting of integration time delay and integration time.The saved data can be imported directly into the Specline analytical software package (see section software). 

Sample chamber Specifications

 
 
AvaLIBS-SC-1
AvaLIBS-SC-2
AvaLIBS-SC-3
Axis, travel path
1 axis, 20mm
2 axis, 20 mm each
3 axis, 50 mm each
Approx. dimensions
100 x 100 x 200 mm
100 x 100 x 250 mm
260 x 280 x 320 mm
Safety protection
Electrical interlock to laser power supply, laser protective window, OD 6+

 


AvaLIBS Modules – an alternative method of configuring a LIBS system

 

AvaLIBS modules offer a convenient and flexible method of configuring a LIBS system.  Six channel (AvaLIBS-module-6) and eight channel (AvaLIBS-module-8) versions are available.  Each optical channel is connected to an individual spectrometer using a single-core optical fiber.  This allows the user to select one or more spectrometers according to their needs.  In addition to the 50 mJ and 100 mJ laser described below, the AvaLIBS modules may be used with a variety of other lasers using an appropriate adaptor plate (contact us for further details).

 libs-8_module_set-up_23-sep-09.jpg
  Example configuration:  AvaLIBS-module-8, 100 mJ Ultra laser, AvaLIBS-SC-3 sample chamber, 8 AvaSpec 2048-LIBS spectrometers

  

 

Configuration table

 

AvaLIBS Module configurations
50 mJ1-6 channels
100 mJ1-6 channels
50 mJ1-8 channels
100 mJ1-8 channels
AvaLIBS Module
AvaLIBS-module-6
AvaLIBS-module-8
Laser
YAG-LASER-50
YAG-LASER-100
YAG-LASER-50
YAG-LASER-100
Sample chambers
Fully compatible with AvaLIBS-SC-1, AvaLIBS-SC-2 and AvaLIBS-SC-3 modular sample chambers
Fully compatible with AvaLIBS-SC-1, AvaLIBS-SC-2 and AvaLIBS-SC-3 modular sample chambers
Fiberoptics
N* x FC-UV200-2
N* x FC-UV200-2
Spectrometers 
Up to 6 spectrometers - as specified by user
Up to 8 spectrometers - as specified by user
Cable set
 IC-LIBS-SET-N*
 IC-LIBS-SET-N*
N* = number of channels

 


 

Ordering Information

 

  AvaLIBS-50 and AvaLIBS-100
  AvaLIBS-BASE   AvaLIBS-50 / AvaLIBS-100 head and spectrometer console, including 6 spectrometers, incl. AvaSoft-FULL, excl. laser
  YAG-LASER-50   50mJ Nd:YAG laser, including power supply
  YAG-LASER-100   100mJ Nd:YAG laser, including power supply
  Imaging kit (optional)   Miniature CCD colour camera, camera mount, IR blocking filter, plug-in power adaptor, 5.6” LCD colour monitor, USB 2.0 video grabber
  AvaLIBS-SC-1   1-axis sample chamber, couples to AvaLIBS head
  AvaLIBS-SC-2   2-axis sample chamber, couples to AvaLIBS head
  AvaLIBS-SC-3   3-axis sample chamber, couples to AvaLIBS head
  Vertical support stand (optional)   Adjustable height vertical support stand for AvaLIBS head
  AvaLIBS-Specline-A (optional)   Spectroscopy analytical software for LIBS, finding and identifying spectral lines, incl. database for atoms and ions
 
 AvaLIBS Modules
 AvaLIBS-module-6  Six-channel LIBS module, including laser beam expander, plasma light collection optics (6 channels), requires a single-core optical fiber cable to connect each channel to a spectrometer (i.e. 6 fibers required for 6 spectrometers)
 Adaptor plate for Ultra laser  For attaching AvaLIBS modules to Ultra laser head (50mJ or 100mJ)
 Imaging kit (optional)  Miniature CCD colour camera, camera mount, IR blocking filter, plug-in power adaptor, 5.6” LCD colour monitor, USB 2.0 video grabber
 Vertical support stand (optional)  Adjustable height vertical support stand for AvaLIBS-module-6 and AvaLIBS-module-8
 FC-UV200-2  Fiber Cable, UV/VIS, 2m, SMA terminated (6 or 8 pcs needed)
 AvaLIBS-module-8  Eight-channel LIBS module, including laser beam expander, plasma light collection optics (6 channels), requires a single-core optical fiber cable to connect each channel to a spectrometer (I.e. 8 fibers required for 8 spectrometers)
 AvaSpec-2048-LIBS AvaLIBS Spectrometer channel, Avaspec-2048-USB2, 2048 CCD detector, specify grating, wavelength range and options, AVASOFT-FULL and LIBS