User Login






Lost Password?
No account yet? Register
Please note:
For ALL software downloads you need to register.

Application Icons

Home arrow Products arrow Spectrometers arrow AvaSpec-NIR256-2.0/2.2/2.5 NIR Fiber Optic Spectrometers





AvaSpec-NIR256-2.0/2.2/2.5 NIR Fiber Optic Spectrometers



The AvaSpec-NIR256-2.0/2.2/2.5 Fiber Optic Spectrometers are based on the AvaBench-50 symmetrical Czerny-Turner design with 256 pixel TE Cooled extended InGaAs Detector Arrays. The spectrometer has a fiber optic entrance connector (standard SMA, others possible), collimating and focusing mirror and diffractional grating. A choice of 4 gratings with different dispersion and blaze angles enables applications in the 900-2500nm range. The AvaSpec-NIR256-2.0/2.2/2.5 includes a 16 bit AD converter, and USB2.0 high speed interface. The AvaSpec- NIR256-2.0/2.2/2.5 have a 2-stage Thermo-electrical Peltier-cooled InGaAs detector, specially suitable for measuring in the NIR range from 1000-2000/2200/2500nm. The newly added AvaSpec-NIR256-2.0 has a very low dark noise level and therefore a much longer integration time can be used for low light level applications.  Digital IO ports enable external triggering and control of shutter and pulsed light sources from the Avantes line of instruments.

The AvaSpec-NIR256 series have a USB2 interface with fast datasampling of 940 spectra per second and datatransfer in 1.56 ms  and supports analog in- and outputs as well. Optional BluetoothÒ (-BT) communication and an SD card for on-board saving of spectra can be added. The AvaSpec-NIR-256-2.0/2.2/2.5 come in a desktop housing and run on 100-240VAC. They come with AvaSoft-basic, a complete manual and USB interface cable. Multiple (up to 127) USB2 spectrometers with different detector types can be externally coupled (see section multi-channel spectrometers).

 


Technical data

Spectrometer platform AvaSpec-NIR256-2.0 AvaSpec-NIR256-2.2 AvaSpec-NIR256-2.5
Optical Bench Symmetrical Czerny-Turner, 50 mm focal length
Wavelength range 1000 - 2000nm 1000 - 2200nm 1000 - 2500nm
Resolution 6.0 - 60 nm 6.0 - 60 nm 6.0-90 nm
Stray light < 0.1%
Sensitivity (AvaLight-HAL, 8 µm fiber) in counts (16-bit A/D)/µW per ms integration time 250 250 200
Detector InGaAs linear array with 2 stage TE-cooling, 256 pixels
Signal/Noise 1500:1
AD converter 16 bit, 500kHzHz
Integration time* 0.52 ms- 6 seconds 0.52-350 ms 0.52-100 ms
Interface USB 2.0 high speed, 480 MbpsRS-232, 115.200 bps
Sample speed with on-board averaging 1.06 ms /scan
Data transfer speed 1.56 msec /scan (USB2)60 ms / scan (RS-232)
Digital IO HD-26 connector, 2 Analog in, 2 Analog out, 3 Digital in, 12 Digital out, trigger, synchronization
Power supply 100-240 VAC, 50/60Hz, 35 W
Dimensions, weight 315 x 235 x 135 mm, 5.1 kg

 *Integrating time with TE cooling set to -20°C 

Grating selection table for AvaSpec-NIR256-2.0/2.2/2.5
Use
Useable range 
Spectral range (nm)
Lines/mm
Blaze (nm)
Order code
NIR
1000-2500
1470
100
2500
NIR100-2.5
NIR
1000-2200
980
150
2000
NIR150-2.0
NIR
900-1750
750
200
1500
NIR200-1.5
NIR
1700-2500
650
200
2600
NIR200-2.6

   

Resolution table (FWHM) for AvaSpec-NIR256-2.0/2.2/2.5

Slit size (µm)

Grating (lines/mm) 50 100 200 500
100 15 25 50 90
150 10 15 30 60
200 6-8 12 24 50
 


 

Ordering Information

 

 AvaSpec-NIR256-2.0  Fiber Optic Spectrometer, 50 mm Avabench, 256 pixel InGaAs detector with 2stage TEC, USB powered high speed USB2 interface, incl. AvaSoft-Basic, USB interface cable, specify grating and wavelength range, OSF-1000, specify slit 
 AvaSpec-NIR256-2.2  Fiber Optic Spectrometer, 50 mm Avabench, 256 pixel InGaAs detector with 2stage TEC, high speed USB2 interface, incl. AvaSoft-Basic, USB interface cable, specify grating and wavelength range, OSF-1000, specify slit
 AvaSpec-NIR256-2.5   Fiber Optic Spectrometer, 50 mm Avabench, 256 pixel InGaAs detector with 2stage TEC, high speed USB2 interface, incl. AvaSoft-Basic, USB interface cable, specify grating and wavelength range, OSF-1000, specify slit
 Options  
 BT    Bluetooth® interface for USB2 platform only, including antenna
 SDXXX    Internal XXX MB SD card for on-board data saving, for USB2 platform only
 SLIT-XX        Slit size, please specify XX = 50, 100, 200, 500 µm

 

                                                                                         







more categories

UV / VIS Spectrometers

NIR spectrometers

OEM spectrometers