The AvaSpec-ULS2048x16/64 Fiber Optic Spectrometers are the initial two new products launched in the Sensline High Sensitivity spectrometer family with high quantum efficiency and high UV and near infrared sensitivity. The detector pixels in the AvaSpec-ULS2048X16 measure 224µm high and for the AvaSpec-ULS2048X64 they measure 896µm high The instrument’s optical design is based on the AvaBench-75ULS Ultra Low Straylight symmetrical Czerny-Turner design. The image sensor is used as a linear array of 2048 pixels binning the vertical 16 or 64 pixels for optimal efficiency. The spectrometer has a fiber optic entrance connector (standard SMA, other options available), collimating and focusing mirrors and is available with choice of 16 different standard diffraction gratings to enable applications in the 200-1160nm range. The AvaSpec-ULS2048x16/64 come with a 16 bit AD converter, and USB2.0 high speed interface. These instruments are especially suitable for measuring low light, fluorescence and UV- applications but also have superior NIR sensitivity relative to front illuminated CCDs. Digital IO ports enable external triggering and control of shutters and pulsed light sources from the Avantes line of instruments. The new AvaSpec-ULS2048x16/64 have detectors with higher speed, lower noise, better anti-blooming and reduced etaloning compared to the AvaSpec-2048x14. Optional Bluetooth® (-BT) communication and an SD card for on-board storage of spectra are available. The AvaSpec-ULS2048x16/64 runs on USB power and comes with AvaSoft-basic, a complete manual and USB interface cable. Multiple USB2 spectrometers with different detector types can be externally coupled (see section multi-channel spectrometers).


Technical Data
Spectrometer Type  AvaSpec-ULS 2048x16-USB2  AvaSpec-ULS 2048x64-USB2
Optical Bench
 
Symmetrical Czerny-Turner, 75 mm focal length
Wavelength range
 
200 - 1160nm
Resolution
 
0.06 -20 nm, depending on configuration (see table)
Stray light
 < 0.2%
< 0.5%
Sensitivity (counts/µW per ms int. time)
 16.000 counts
32.000 counts
UV Quantum efficiency
 
60% (200-300nm)
Detector
 Back-thinned CCD image sensor 2048x16 pixels
Back-thinned CCD image sensor 2048x64 pixels
Signal/Noise
 
500:1
AD converter
 
16 bit, 1.3 MHz
Integration time
 1.82 ms - 60 sec
2.4 ms - 60 sec
Interface
 
USB 2.0 high speed, 480 Mbps
RS-232, 115.200 bps
Sample speed with on-board averaging
 1.82 ms / scan
2.24 ms /scan
Data transfer speed

 1.82 ms / scan (USB2)
 432 ms / scan (RS-232)

2.24 ms / scan (USB2)
432 ms / scan (RS-232)
Digital IO
 
HD-26 connector, 2 Analog in, 2 Analog out, 3 Digital in, 12 Digital out, trigger, synchronization
Power supply
 
Default USB power, 450 mA
Or with SPU2 external 12VDC, 450 mA
Dimensions, weight
 
175 x 110 x 44 mm(1 channel), 855 grams

Grating selection table for AvaSpec-2048x16 / 64 

Use
Useable range 
Spectral range (nm)
Lines/mm
Blaze (nm)
Order code
UV/VIS/NIR
200-1160**
960**
300
300
UA
UV/VIS
200-850
520
600
300
UB
UV
200-750
250-220*
1200
250
UC
UV
200-650
165-145*
1800
UV
UD
UV
200-580
115-70*
2400
UV
UE
UV
220-400
70-45*
3600
UV
UF
UV/VIS
250-850
520
600
400
BB
VIS/NIR
300-1160**
800**
300
500
VA
VIS
360-1000
500
600
500
VB
VIS
300-800
250-200*
1200
500
VC
VIS
350-750
145-100*
1800
500
VD

VIS

350-640
75-50*
2400
VIS
VE
NIR
500-1050
500
600
750
NB
NIR
500-1050
220-150*
1200
750
NC
NIR
600-1160**
500**
300
1000
IA
NIR
600-1160
500
600
1000
IB

* depends on the starting wavelength of the grating; the higher the wavelength, the bigger the dispersion and the smaller the range to select.
**  please note that not all 2048 pixels will be used for the useable range  

Resolution table (FWHM) for AvaSpec-2048x16 / 64

Slit size (µm)
Grating (lines/mm)
10
25
50
100
200
500
300
1.0
1.5
2.5
4.7
9.2
21.3
600
0.5
0.7
1.2
2.4
4.6
10.8
1200
0.16-0.36*
0.38
0.6
1.1
2.3
5.4
1800
0.09-0.18*
0.15-0.24*
0.4
0.8
1.6
3.6
2400
0.08-0.12*
0.12-0.18*
0.3
0.6
1.1
2.7
3600
0.06
0.08-0.10*
0.2
0.4
0.8
1.8


 

 

 

 


*depends on the starting wavelength of the grating; the higher the wavelength, the bigger the dispersion and the better the resolution


Ordering Information

AvaSpec-ULS2048x16-USB2  Ultra low Straylight Fiber Optic Spectrometer, 75 mm Avabench, 2048x16 pixel back-thinned CCD detector, USB powered high speed USB2 interface, incl. AvaSoft-Basic, USB interface cable, specify grating, wavelength range and options
AvaSpec-ULS2048x64-USB2  Ultra low Straylight Fiber Optic Spectrometer, 75 mm Avabench, 2048x64 pixel back-thinned CCD detector, USB powered high speed USB2 interface, incl. AvaSoft-Basic, USB interface cable, specify grating, wavelength range and options

Options

SPU2  incl. switch for USB power or external power for RS-232/BT
SPU2-BT Bluetooth® interface for USB2 platform only, including antenna and switch  
SDXXX Internal XXX MB SD card for on-board data saving
SLIT-XX   Slit size, please specify XX = 10, 25, 50, 100, 200, 500 µm
OSF-YYY   Order sorting filter for reduction of 2nd order effects, 1 mm thick, please specify -  YYY= 305, 385, 475, 515, 550, 600 nm
 OSC  Order sorting coating with 590nm long pass filter for VA, BB (>350nm) and VB gratings in AvaSpec-2048x16/64, recommended with OSF-305
OSC-UA  Order sorting coating with 350 and 590nm longpass filter for UA gratings in AvaSpec-2048x16/64
OSC-UB  Order sorting coating with 350 and 590nm longpass filter for UB or BB (<350nm) gratings in AvaSpec-2048x16/64

>>>> Click HERE for a pdf of the datasheet

 







more categories

StarLine Spectrometers

NIRLine spectrometers

OEM spectrometers

SensLine spectrometers

Spectrometer Systems

Spectrometer accessories