The AvaSpec-ULS2048x16/64 Fiber Optic Spectrometers are the initial two new products launched in the Sensline High Sensitivity spectrometer family with high quantum efficiency and high UV and near infrared sensitivity. The detector pixels in the AvaSpec-ULS2048X16 measure 224µm high and for the AvaSpec-ULS2048X64 they measure 896µm high The instrument’s optical design is based on the AvaBench-75ULS Ultra Low Straylight symmetrical Czerny-Turner design. The image sensor is used as a linear array of 2048 pixels binning the vertical 16 or 64 pixels for optimal efficiency. The spectrometer has a fiber optic entrance connector (standard SMA, other options available), collimating and focusing mirrors and is available with choice of 16 different standard diffraction gratings to enable applications in the 200-1160nm range. The AvaSpec-ULS2048x16/64 come with a 16 bit AD converter, and USB2.0 high speed interface. These instruments are especially suitable for measuring low light, fluorescence and UV- applications but also have superior NIR sensitivity relative to front illuminated CCDs. Digital IO ports enable external triggering and control of shutters and pulsed light sources from the Avantes line of instruments. The new AvaSpec-ULS2048x16/64 have detectors with higher speed, lower noise, better anti-blooming and reduced etaloning compared to the AvaSpec-2048x14. Optional Bluetooth® (-BT) communication and an SD card for on-board storage of spectra are available. The AvaSpec-ULS2048x16/64 runs on USB power and comes with AvaSoft-basic, a complete manual and USB interface cable. Multiple USB2 spectrometers with different detector types can be externally coupled (see section multi-channel spectrometers).
Technical Data
| Spectrometer Type |
AvaSpec-ULS 2048x16-USB2 |
AvaSpec-ULS 2048x64-USB2 |
|
Optical Bench
|
Symmetrical Czerny-Turner, 75 mm focal length
|
|
Wavelength range
|
200 - 1160nm
|
|
Resolution
|
0.06 -20 nm, depending on configuration (see table)
|
|
Stray light
|
< 0.2% |
< 0.5%
|
|
Sensitivity (counts/µW per ms int. time)
|
16.000 counts |
32.000 counts
|
|
UV Quantum efficiency
|
60% (200-300nm)
|
|
Detector
|
Back-thinned CCD image sensor 2048x16 pixels |
Back-thinned CCD image sensor 2048x64 pixels
|
|
Signal/Noise
|
500:1
|
|
AD converter
|
16 bit, 1.3 MHz
|
|
Integration time
|
1.82 ms - 60 sec |
2.4 ms - 60 sec
|
|
Interface
|
USB 2.0 high speed, 480 Mbps
RS-232, 115.200 bps
|
|
Sample speed with on-board averaging
|
1.82 ms / scan |
2.24 ms /scan
|
|
Data transfer speed
|
1.82 ms / scan (USB2)
432 ms / scan (RS-232)
|
2.24 ms / scan (USB2)
432 ms / scan (RS-232)
|
|
Digital IO
|
HD-26 connector, 2 Analog in, 2 Analog out, 3 Digital in, 12 Digital out, trigger, synchronization
|
|
Power supply
|
Default USB power, 450 mA
Or with SPU2 external 12VDC, 450 mA
|
|
Dimensions, weight
|
175 x 110 x 44 mm(1 channel), 855 grams
|
Grating selection table for AvaSpec-2048x16 / 64
|
Use
|
Useable range
|
Spectral range (nm)
|
Lines/mm
|
Blaze (nm)
|
Order code
|
|
UV/VIS/NIR
|
200-1160**
|
960**
|
300
|
300
|
UA
|
|
UV/VIS
|
200-850
|
520
|
600
|
300
|
UB
|
|
UV
|
200-750
|
250-220*
|
1200
|
250
|
UC
|
|
UV
|
200-650
|
165-145*
|
1800
|
UV
|
UD
|
|
UV
|
200-580
|
115-70*
|
2400
|
UV
|
UE
|
|
UV
|
220-400
|
70-45*
|
3600
|
UV
|
UF
|
|
UV/VIS
|
250-850
|
520
|
600
|
400
|
BB
|
|
VIS/NIR
|
300-1160**
|
800**
|
300
|
500
|
VA
|
|
VIS
|
360-1000
|
500
|
600
|
500
|
VB
|
|
VIS
|
300-800
|
250-200*
|
1200
|
500
|
VC
|
|
VIS
|
350-750
|
145-100*
|
1800
|
500
|
VD
|
VIS
|
350-640
|
75-50*
|
2400
|
VIS
|
VE
|
|
NIR
|
500-1050
|
500
|
600
|
750
|
NB
|
|
NIR
|
500-1050
|
220-150*
|
1200
|
750
|
NC
|
|
NIR
|
600-1160**
|
500**
|
300
|
1000
|
IA
|
|
NIR
|
600-1160
|
500
|
600
|
1000
|
IB
|
* depends on the starting wavelength of the grating; the higher the wavelength, the bigger the dispersion and the smaller the range to select.
** please note that not all 2048 pixels will be used for the useable range
Resolution table (FWHM) for AvaSpec-2048x16 / 64
|
Slit size (µm)
|
|
Grating (lines/mm)
|
10
|
25
|
50
|
100
|
200
|
500
|
|
300
|
1.0
|
1.5
|
2.5
|
4.7
|
9.2
|
21.3
|
|
600
|
0.5
|
0.7
|
1.2
|
2.4
|
4.6
|
10.8
|
|
1200
|
0.16-0.36*
|
0.38
|
0.6
|
1.1
|
2.3
|
5.4
|
|
1800
|
0.09-0.18*
|
0.15-0.24*
|
0.4
|
0.8
|
1.6
|
3.6
|
|
2400
|
0.08-0.12*
|
0.12-0.18*
|
0.3
|
0.6
|
1.1
|
2.7
|
|
3600
|
0.06
|
0.08-0.10*
|
0.2
|
0.4
|
0.8
|
1.8
|
*depends on the starting wavelength of the grating; the higher the wavelength, the bigger the dispersion and the better the resolution
Ordering Information
| AvaSpec-ULS2048x16-USB2 |
Ultra low Straylight Fiber Optic Spectrometer, 75 mm Avabench, 2048x16 pixel back-thinned CCD detector, USB powered high speed USB2 interface, incl. AvaSoft-Basic, USB interface cable, specify grating, wavelength range and options |
| AvaSpec-ULS2048x64-USB2 |
Ultra low Straylight Fiber Optic Spectrometer, 75 mm Avabench, 2048x64 pixel back-thinned CCD detector, USB powered high speed USB2 interface, incl. AvaSoft-Basic, USB interface cable, specify grating, wavelength range and options |
Options
| SPU2 |
incl. switch for USB power or external power for RS-232/BT |
| SPU2-BT |
Bluetooth® interface for USB2 platform only, including antenna and switch |
| SDXXX |
Internal XXX MB SD card for on-board data saving |
| SLIT-XX |
Slit size, please specify XX = 10, 25, 50, 100, 200, 500 µm |
| OSF-YYY |
Order sorting filter for reduction of 2nd order effects, 1 mm thick, please specify - YYY= 305, 385, 475, 515, 550, 600 nm |
| OSC |
Order sorting coating with 590nm long pass filter for VA, BB (>350nm) and VB gratings in AvaSpec-2048x16/64, recommended with OSF-305 |
| OSC-UA |
Order sorting coating with 350 and 590nm longpass filter for UA gratings in AvaSpec-2048x16/64 |
| OSC-UB |
Order sorting coating with 350 and 590nm longpass filter for UB or BB (<350nm) gratings in AvaSpec-2048x16/64 |
>>>> Click HERE for a pdf of the datasheet
|