AvaSpec-2048L Fiber Optic Spectrometer with LARGE pixels



The new AvaSpec-2048L Fiber Optic Spectrometer has been recently added to the Avantes product line and has many technical specifications comparable to the standard AvaSpec-2048.
The difference is the CCD detector, the AvaSpec-2048L uses the Sony ILX511 detector whereas the AvaSpec-2048 uses the ILX554B detector.
The AvaSpec-2048L is very useful for applications where larger fiber entrance diameter needs to be projected on the larger CCD detector pixels of 200µm height.
The AvaSpec-2048L also has a better signal to noise performance and shows no peak inversion and dark increase at the end of the detector array at longer integration times. The FWHM resolution of the 2048L is ca. 20% less good than for the standard 2048.Non-linearity is also a little less, so a NL-calibration per channel is recommended.
All options for the AvaSpec-2048 are also available on the AvaSpec-2048L, the DCL-UV/VIS-200 has been added to optimize sensitivity for fiber diameters larger than 200µm.


Technical Data

 

Spectrometer
AvaSpec-2048-USB2
AvaSpec-2048L-USB2
Detector
Sony ILX-554B
Sony ILX-511
Pixel size
14 X 56 µm
14 x 200µm
Sensitivity (V/lx.s)
240
200
Sensitivity (AvaLight-HAL, 8 µm fiber) counts/µW  per ms integration time
20,000
20,000
Peak Wavelength
550 nm
450 nm
Signal/Noise
200:1
300:1
Dark Noise
40
24
Ext trigger delay
1.3µs
3.0µs

 


Ordering Information
 

AvaSpec-2048L-USB2         Fiber Optic Spectrometer, 75 mm Avabench, 2048 pixel 14x200µm CCD detector, USB powered high speed USB2 interface, incl. AvaSoft-Basic, USB interface cable, specify grating, wavelength range and options 
 Options    All options for the AvaSpec-2048-USB2     
 extra Options:
 
 DCL-UV/VIS-200  Quartz Detector Collection Lens (200-1100nm)