NIR Spectrometers up to 1700nm (TEC Cooled)

NIR Spectrometers up to 1700nm (TEC Cooled)
 
||| |||

For measurements in the near infrared range out to 1.7 µm or 1700 nm, Avantes offers three spectrometer configurations. The AvaSpec-NIR256-1.7 is the most economical in the NIR-range of spectrometers made by Avantes, but offers exceptional performance specifications such as a sample speed of only 1.06 ms/scan and integration times as fast as 10 µs. This instrument is also available in RS-version meaning you can easily change the slit in a matter of seconds.

The AvaSpec-NIR256/512-1.7 TEC spectrometers are also affordably priced and can be operated in two distinct gain-setting modes, the high-sensitivity mode (HS, default) and the low-noise (LN) mode. Setting of the different gain modes can be done via a software option. All NIR-1.7 instruments are available with a choice of four different gratings, making it possible to choose the bandwidth fitting your application. Digital and analog I/O ports enable external triggering and control over the shutter and pulsed light sources from the Avantes line of products. The uncooled AvaSpec-NIR256-1.7 is USB2 powered, while the TEC cooled instruments require external power.

 

 

Technical

Spectrometer AvaSpec-NIR256-1.7TEC AvaSpec-NIR512-1.7TEC
Optical Bench Symmetrical Czerny-Turner, 50 mm focal length Symmetrical Czerny-Turner, 50 mm focal length, TE cooled
Wavelength range 900 - 1750nm
Resolution 3.0-50 nm (400 l/mm grating) 2.5-50 nm (400 l/mm grating)
Stray light < 0.5%
Sensitivity HS in counts/µW per ms integration time 2,700.000 (integral 1000-1750 nm)
Signal/Noise HS 1700:1
Integration time HS 20 µs - 1 second
Sensitivity LN in counts /µW per ms integration time 96,000 (integral 1000-1750 nm)
Signal/Noise LN 3600:1
Integration time LN 20 µs - 35 seconds
Detector TE-cooled InGaAs linear array, 256 pixels, 50 µm x 500 µm TE-cooled InGaAs linear array, 512 pixels, 25 µm x 500 µm
AD converter 16 bit, 2.4 MHz
Interface USB 2.0 high speed, 480 Mbps RS-232, 115.200 bps
Sample speed with on-board averaging 0.19 ms /scan @ 0.08ms int. time 0.31 ms /scan @ 0.09ms int. time
Data transfer speed 1.0 ms /scan (USB2)
60 ms /scan (RS-232)
1.2 ms /scan (USB2)
120 ms /scan (RS-232)
Digital IO HD-26 connector, 2 Analog in, 2 Analog out, 3 Digital in, 12 Digital out, trigger, synchronization
Power supply 100-240 VAC, 50-60 Hz, 35W
Dimensions, weight 244 x 144 x 254 mm, 5.1 Kg

Gratings

  Use Spectral range (nm) Lines/mm  Blaze (nm)  Order code
    256 256/512TEC      
NIR 900-1750 850 650 200 1500 NIR200-1.5
NIR 1000-1700 340 250 400 1600 NIR400-1.6

NIR

900-1400

200 n.a.

600

1200

NIR600-1.2

NIR 1300-1600 170 n.a. 600 1600 NIR600-1.6

Resolution

Slit size (µm)
Grating (lines/mm) 25* 50 100 200 500
200 5 6 12 24 50
400 2.5 3 6 12 25
600 n.a. 2 4 8

18

(*only for AvaSpec-NIR512TEC)

Order Info

AvaSpec-NIR256-1.7TEC Fiber-optic Spectrometer, 50 mm AvaBench, 256 pixel InGaAs detector with 2-stage TE cooling, USB powered high-speed USB2 interface, incl. AvaSoft-Basic, USB interface cable, OSF-850/1000-3.
Specify grating, wavelength range and slit
AvaSpec-NIR512-1.7TEC Fiber-optic Spectrometer, 50 mm AvaBench, 512 pixel InGaAs detector with 2-stage TE cooling, USB powered high-speed USB2 interface, incl. AvaSoft-Basic, USB interface cable, OSF-850/1000-3.
Specify grating, wavelength range and slit

Options

Options  
-RS Replaceable slit for the AvaSpec-NIR256-1.7 only
SLIT-XX-RS Replaceable slit with SMA connector, specify slit size XX=25, 50, 100 or 200 µm, only in combination with the AvaSpec-NIR256-1.7-RS
SLIT-XX-RS-FCPC As SLIT-XX-RS, but with FC/PC connector
SLIT-XX

Slit size, please specify XX = 25*, 50, 100, 200, 500 µm