Combining exceptional quantum efficiency with high speed communication is the value proposition of the AvaSpec-ULS2048XL EVO spectrometer. Unlike many back-thinned CCD spectrometers that have two dimensional arrays, the ULS2048XL-EVO has large monolithic pixels of 14×500 microns with exceptional efficiency in the UV range, from 200 to 400 nm, and the NIR range, from 950 to 1160 nm.
The instrument also has an electronic shutter, which enables integration times as low as 2 microseconds. To further enhance sensitivity, a detector collection lens is available, which improves sensitivity up to 60% when combined with larger core fibers.
Product Information
Optical bench | Ultra-low stray light (ULS) symmetrical Czerny-Turner, 75 mm focal length |
Wavelength range | 200-1160 nm |
Resolution | 0.09-20 nm, depending on configuration |
Stray light | <0.5% |
Sensitivity in counts/µW per ms integration time | 460,000 |
UV quantum efficiency | 60% (200-300 nm) |
Detector | Back-thinned CCD image sensor, 2048 pixels |
Signal/noise | 525:1 |
AD converter | 16 bit, 1 mHz |
Integration time | 2 µs-20 sec |
Interface | USB 3.0 high speed, 5 Gbps, Gigabit Ethernet 1 Gbps |
Sample speed with store to RAM | 2.44 ms/scan |
Readout noise | 9.8 cnts RMS |
Dark noise | 4.5 cnts RMS |
Dynamic range | 13,700 |
Data transfer speed | 2.44 ms/scan (USB3) |
Digital IO | HD-26 connector, 2 Analog in, 2 Analog out, 3 Digital in, 12 Digital out, trigger, synchronization |
Power supply | Default USB power, 700 mA Or external 12VDC, 360 mA |
Dimensions, weight | 175 x 127 x 44,5 mm (1 channel), 1180 g |
Use | Useable range (nm) | Spectral range (nm) | Lines/mm | Blaze (nm) | Order code |
---|---|---|---|---|---|
UV/VIS/NIR | 200-1160** | 960** | 300 | 300 | UA |
UV/VIS/NIR | 200-1100** | 900** | 300 | 300/1000 | UNA-DB |
UV/VIS | 200-850 | 520 | 600 | 300 | UB |
UV | 200-750 | 250-220* | 1200 | 250 | UC |
UV | 200-650 | 165-145* | 1800 | UV | UD |
UV | 200-580 | 115-70* | 2400 | UV | UE |
UV | 200-400 | 70-45* | 3600 | UV | UF |
UV/VIS | 250-850 | 520 | 600 | 400 | BB |
VIS/NIR | 300-1160** | 860** | 300 | 500 | VA |
VIS | 360-1000 | 500 | 600 | 500 | VB |
VIS | 300-800 | 250-200* | 1200 | 500 | VC |
VIS | 350-750 | 145-100* | 1800 | 500 | VD |
VIS | 350-640 | 75-50* | 2400 | VIS | VE |
NIR | 500-1050 | 500 | 600 | 750 | NB |
NIR | 500-1050 | 220-150* | 1200 | 750 | NC |
NIR | 600-1160 | 350-300 | 830 | 800 | SI |
NIR | 600-1160** | 560** | 300 | 1000 | IA |
NIR | 600-1160 | 500 | 600 | 1000 | IB |
Slit size (µm) | 10 | 25 | 50 | 100 | 200 | 500 |
---|---|---|---|---|---|---|
300 lines/mm grating | 1.4 | 1.5 | 2.5 | 4.8 | 9.2 | 21.3 |
600 lines/mm gratings | 0.70-0.80* | 0.75-0.85* | 1.2 | 2.4 | 4.6 | 10.8 |
830 lines/mm grating | 0.42-0.48* | 0.50-0.58* | 0.93 | 1.7 | 3.4 | 8.5 |
1200 lines/mm grating | 0.25-0.31* | 0.37-0.43* | 0.52-0.66* | 1.1 | 2.3 | 5.4 |
1800 lines/mm grating | 0.17-0.21* | 0.26-0.32* | 0.34-0.42* | 0.8 | 1.6 | 3.6 |
2400 lines/mm grating | 0.12-0.18* | 0.18-0.24* | 0.26-0.34* | 0.44-0.64* | 1.1 | 2.7 |
3600 lines/mm grating | 0.09-0.12* | 0.11-0.15* | 0.19 | 0.4 | 0.8 | 1.8 |
AvaSpec-ULS2048XL-EVO | Ultra-low Stray-light Fiber-optic Spectrometer, 75 mm AvaBench, 2048 large 500 µm pixel back-thinned CCD detector, USB powered, high-speed USB3 and ETH interface, incl. AvaSoft-Basic, USB interface cable. Specify grating, wavelength range and options |
PS-12V/1.0A | External power supply, needed for operation in ETH mode |
-RS | Replaceable slit |
DCL-UV/VIS-200 | Detector Collection Lens to enhance sensitivity, Quartz, 200-1100 nm |
SLIT-XX | Slit size, please specify XX = 10, 25, 50, 100, 200 or 500 µm |
SLIT-XX-RS | Replaceable slit with SMA connector , specify slit size XX=25, 50, 100, 200 or 500 µm. Only in combination with AvaSpec-ULS2048XL-EVO-RS |
SLIT-XX-RS-FCPC | As SLIT-XX-RS, but with FC/PC connector |
OSF-YYY | Order-sorting filter for reduction of 2nd order effects, 1 mm thick, please specify YYY= 305, 385, 475, 515, 550 or 600 nm |
OSC | Order-sorting coating with 600 nm long-pass filter for VA, BB (>350 nm) and VB gratings in AvaSpec-2048XL, recommended with OSF-305 |
OSC-UA | Order-sorting coating with 350 and 600 nm linear variable filter for UA gratings in AvaSpec-ULS2048XL |
OSC-UB | Order-sorting coating with 350 and 600 nm long-pass filter for UB or BB (<350 nm) gratings in AvaSpec-ULS2048XL |
-FCPC | FC/PC fiber-optic connector |