Technical
Optical Bench | Symmetrical Czerny-Turner, 75 mm focal length |
Wavelength range | 200-1100 nm |
Resolution | 0.05-20 nm, depending on configuration (see table) |
Stray-light | 0.04-0.1%, depending on the grating |
Sensitivity | 310,000 counts/µW per ms integration time |
Detector | CCD linear array, 2048 pixels |
Signal/Noise | 200:1 |
AD converter | 16 bit, 2 MHz |
Integration time | 1.11 ms - 10 minutes |
Interface | USB 2.0 high speed, 480 Mbps RS-232, 115.200 bps |
Sample speed with on-board averaging | 1.1 ms /scan |
Data transfer speed | 1.8 ms / scan (USB2) 430 ms / scan (RS-232) |
Digital IO | HD-26 connector, 2 Analog in, 2 Analog out, 3 Digital in, 12 Digital out, trigger, sync. |
Power supply | Default USB power, 350 mA or with SPU2 external 12VDC, 150 mA |
Dimensions, weight | 175 x 110 x 44 mm (1 channel), 716 grams |
Gratings
Use | Useable range | Spectral range (nm) | Lines/mm | Blaze (nm) | Order code |
UV/VIS/NIR | 200-1100** | 900** | 300 | 300 | UA |
UV/VIS/NIR | 200-1100** | 900** | 300 | 300/1000 | UNA-DB |
UV/VIS | 200-850 | 520 | 600 | 300 | UB |
UV | 200-750 | 250-220* | 1200 | 250 | UC |
UV | 200-650 | 165-145* | 1800 | UV | UD |
UV | 200-580 | 115-70* | 2400 | UV | UE |
UV | 200-400 | 70-45* | 3600 | UV | UF |
UV/VIS | 250-850 | 520 | 600 | 400 | BB |
VIS/NIR | 300-1100** | 800** | 300 | 500 | VA |
VIS | 360-1000 | 500 | 600 | 500 | VB |
VIS | 300-800 | 250-200* | 1200 | 500 | VC |
VIS | 350-750 | 145-90* | 1800 | 500 | VD |
VIS | 350-640 | 75-50* | 2400 | VIS | VE |
NIR | 500-1050 | 500 | 600 | 750 | NB |
NIR | 500-1050 | 220-150* | 1200 | 750 | NC |
NIR | 600-1160 | 350-300* | 830 | 800 | SI |
NIR | 600-1100** | 500** | 300 | 1000 | IA |
NIR | 600-1100 | 500 | 600 | 1000 | IB |
*depends on the starting wavelength of the grating; the higher the wavelength, the bigger the dispersion and the smaller the range to select.
**please note that not all 2048 pixels will be used for the useable range
Resolution
Slit size (µm) | ||||||
Grating (lines/mm) | 10 | 25 | 50 | 100 | 200 | 500 |
300 | 0.80-0.90* | 1.10-1.20* | 2.30 | 4.60 | 9.00 | 20.0 |
600 | 0.40-0.50* | 0.63 | 1.15 | 2.31 | 4.50 | 10.0 |
830 | 0.28 | 0.40 | 0.80 | 1.60 | 3.20 | 8.0 |
1200 | 0.18-0.22* | 0.29 | 0.61 | 1.18 | 2.20 | 5.4 |
1800 | 0.10-0.16* | 0.19 | 0.35-0.42* | 0.80 | 1.60 | 3.6 |
2400 | 0.08-0.11* | 0.10-0.15* | 0.28 | 0.55 | 1.10 | 2.7 |
3600 | 0.05-0.08* | 0.10 | 0.18 | 0.38 | 0.75 | 1.8 |
* depends on the starting wavelength of the grating; the higher the wavelength, the bigger the dispersion and the smaller the range to select.
Order Info
AvaSpec-ULS2048-USB2 | Fiber-optic Spectrometer, 75 mm AvaBench, 2048 pixel CCD detector, USB powered, high-speed USB2 interface, incl. AvaSoft-Basic, USB interface cable. Specify grating, wavelength range and options |
Options
-SPU2 | Incl. switch for USB powered USB2 or external power for RS-232 |
-RS | Replaceable slit |
DUV | Deep UV detector coating >150 nm |
DCL-UV/VIS | Detector collection Lens to enhance sensitivity, Quartz, 200-1100nm |
SLIT-XX |
Slit size, please specify XX = 10, 25, 50, 100, 200, 500 µm |
SLIT-XX-RS | Replaceable slit with SMA connector, specificy slit size XX=25, 50, 100 or 200 µm. Only in combination with AvaSpec-ULS2048-USB2-RS |
SLIT-XX-RS-FCPC | As SLIT-XX-RS, but with FC/PC connector |
OSF-YYY-3 | Order sorting filter for reduction of 2nd order effects, 3 mm thick, please specify YYY= 385, 475, 515, 550, 600 nm |
OSC | Order sorting coating with 600 nm long pass filter for VA, BB (>350nm) and VB gratings in AvaSpec-ULS2048 |
OSC-UA | Order sorting coating with 350 and 600 nm linear variable filter for UA gratings in AvaSpec-ULS2048 |
OSC-UB | Order sorting coating with 350 and 600 nm longpass filter for UB or BB (<350nm) gratings in AvaSpec-ULS2048 |
-FCPC | FC/PC fiber-optic connector |
Downloads
folder
AvaSoft
spectroscopy software